Digital Leadset

Connecting/Disconnecting the Leadset
The digital leadset connects to the Mixed Signal interface on the front of the instrument.
To connect the leadset to the instrument, push the connector into the Mixed Signal interface below
the front panel until you hear a click.
To remove the leadset, press and hold the buttons on each side of the connector, then pull out to
release.
Grounding Leads
Each flying lead has a signal and a ground connection. A variety of ground extenders and flying ground
leads are available for different probing needs.
To achieve optimal signal integrity, connect the ground at the tip of the flying lead for each input used
in your measurements. Use either the provided ground extenders or ground flying leads to make the
ground connection.
The digital leadset provided with -MS model
oscilloscopes enables input of up-to-16 lines of
digital data. Lines can be organized into four
logical groups and renamed appropriately.
The digital leadset features two digital banks
with separate Threshold controls, making it
possible to simultaneously view data from
different logic families.
Oscilloscope Overview