6.13 Module Specifications

Acceleration/Voltage Module (with AAF) (701275)

Item
Standard operating conditions
Effective measurement range
Number of input channels
Input coupling
Maximum sample rate
Input format
Frequency characteristics
(−3 dB point when a sine wave
of amplitude ±3 div is applied)
Voltage-axis sensitivity setting
Maximum input voltage
(at a frequency of 1 kHz or less)
Maximum allowable common
mode voltage
(at a frequency of 1 kHz or less)
Vertical (voltage) axis accuracy
1
DC accuracy
Input connector
Input impedance
−3 dB point when AC coupled
low frequency attenuation point
Common mode rejection ratio
Residual noise level
(Input section shorted)
A/D conversion resolution
Temperature coefficient
Bandwidth limit
Probe attenuation setting
Compatible probes/cables
Sensor supply current (voltage)
Applicable acceleration sensor
6-36
Specifications
Temperature:
23 ±5°C
Humidity:
20 to 80% RH
After a 30-minute warm-up and after calibration
20 div (display range: 10 div)
2
AC, DC, GND, ACCL (acceleration), and GND
100 kS/s
Isolated unbalanced
1
Waveform observation mode:
Acceleration measurement mode:
5 mV/div to 10 V/div (1-2-5 steps) (when using 1:1 probe attenuation)
Acceleration (±5 V = ×1 range): ×0.1 to ×1 to ×100 (in 1-2-5 steps)
42 V (DC+Acpeak)
Working voltage of safety standard
42 V (DC+ACpeak) (CAT II, 30 Vrms)
Waveform measurement mode DC accuracy:
Acceleration measurement mode:
Metal BNC connector
1 MΩ±1%, approx. 35 pF
0.4 Hz or less (0.04 Hz or less when using the 701940) (typical value
80 dB (50/60 Hz) or more (typical value
±100 µV or ±0.01 div, whichever is greater (typical value
16 bits (2400LSB/div)
When in waveform measurement mode (excluding AUTO filter)
Zero point:
±(0.02% of 10 div)/°C (typical value
Gain:
±(0.02% of 10 div)/°C (typical value
Select from Full, Auto, 4 kHz, 400 Hz, and 40 Hz
Cutoff characteristics:
Cutoff frequency (fc) when set to AUTO
Sample rate of 100 kHz or higher:
Sample rate of 100 Hz to 50 kHz:
Sample rate of 50 Hz or less:
Cutoff characteristics when set to AUTO: −65 dB at 2×fc (typical value
Voltage probe 1:1, 10:1, 100:1, or 1000:1
Current probe 1 A:1 V, 10 A:1 V (for the 701932/701933), 100 A:1 V (for the 701930/701931)
Connection cable (for low voltage 1:1)
366926 (non-isolated type BNC-alligator clip × 2: For measuring low voltage less than or
equal to 42 V (DC+ACpeak))
Voltage probe (10:1 passive probe)
701940 17 to 46 pF: For measuring 600 V (DC+ACpeak) or less
Current probe (power can be supplied from the DL850E/DL850EV)
701930 (150 A), 701931 (500 A), 701932 (30 A), 701933 (30 A)
OFF/4 mA ± 10% (approx. 22 VDC)
Built-in amplifier type
Kistler Instrument Corporation: Piezotron, PCB Piezotronics Incorporated: ICP, ENDEVCO
Corporation: ISOTRON, etc.
DC to 40 kHz
0.4 Hz to 40 kHz
2
3
4
)
−12 dB/OCT (typical value
fc = 40 kHz
fc = 40% of the sampling rate
fc = 20 Hz
±(0.25% of 10 div)
±(0.5% of 10 div) at 1 kHz
4
)
4
)
4
)
4
)
4
, excluding AUTO)
4
)
IM DL850E-03EN