5.5 GO/NO-GO Determination I/O and External Start/Stop Input (EXT I/O)
Output Signal
NOGO OUT (Negative logic)
When the determination result is NO-GO (fail), the output signal level temporarily changes from high
(H) to low (L).
GO OUT (Negative logic)
When the determination result is GO, the output signal level temporarily changes from high (H) to
low (L).
Signal Output Circuit
GO/NO-GO I/O Timing
START IN
NOGO OUT
GO OUT
START IN is not referenced when Remote is
set to OFF; GO OUT and NO-GO OUT are
output according to internal GO/NO-GO
determination.
5-8
5 V
GO OUT
4.7 kΩ
NO-GO OUT
120 Ω
100 pF
100 ms or more
50 ms
or more
Acquisition
Judgment
Result
0 ms
or more
Acquisition
Judgment
The signal remains low until the DL850E/DL850EV
is ready to accept the next measurement.
The duration is longer during output to a built-in
printer or internal storage medium.
Result
IM DL850E-03EN