External Start/Stop Input (EXT I/O)
You can use an external signal to start and stop the DL850E/DL850EV.
External Start/Stop Input Terminal
The External Start/Stop Input terminal and the GO/NO-GO I/O terminal are the same terminal. This
terminal is used as an external start/stop input when the GO/NO-GO determination I/O function is not
used (when on the GO/NO-GO menu, Mode is set to OFF).
Specifications
The connector uses an RJ-11 modular jack. Connect the separately sold 720911 cable to the
connector.
EXT I/O
6
1
Connector on
the DL850E/DL850EV
Input level: TTL (0 to 5 V)
Circuit Diagram for External Start/Stop Input
120 Ω
2
(pin no.)
• Contact input is possible
2
5
Note
Low and high edges are used to detect starts and stops.
You can select whether to enable high edges (stops) in the external start/stop signal or disable. For
information about how to do so, see section 18.8, "Configuring the Environment Settings," in the user's
manual.
IM DL850E-03EN
5.5 GO/NO-GO Determination I/O and External Start/Stop Input (EXT I/O)
Pin no.
Signal
1
EXT EVENT IN
2
START IN
3
NC (no connection)
4
NC (no connection)
5
GND
6
NC (no connection)
+5 V
4.7 kΩ
LCX14
100 pF
+5 V
4.7 kΩ
LCX14
120 Ω
100 pF
Close to start,
open to stop
Manual event. Event input occurs on low edge.
Starts on low edge, stops on high
High
Low
START
App
STOP