3.5

Synchronization with the Controller

Achieving Synchronization
IM AQ7277-17EN
If you send the following program message when receiving averaging measurement
data, the data may be retrieved before the completion of the measurement.
ACQuire: AVERage: STARt; : WAVedata: DISPlay: SEND: ASCii?
In such case, the following method must be used to synchronize with the end of the
acquisition.
Using the STATus:CONDition? Query
The "STATus:CONDition?" command is used to query the contents of the condition
register.
Whether averaging measurement waveforms are being retrieved can be determined by
reading bit 1 of the condition register.
If bit 1 of the condition register is "1", waveforms are being retrieved. Otherwise, it is
stopped.(Bit 1 is "0".)
Note
For details on the condition register, see section 4.4, "Condition Register."