6.12 General Specifications; Index - YOKOGAWA DL850E User Manual

Scopecorder
Hide thumbs Also See for DL850E:
Table of Contents

Advertisement

6.12 General Specifications

Item
Standard operating
conditions
Recommended calibration
period
Warm-up time
Storage environment
Operating environment
Standard model
Rated supply voltage
Permitted supply voltage
range
Rated supply frequency
Permitted supply voltage
frequency range
Power fuse
Maximum power
consumption
Withstand voltage
Insulation resistance
External dimensions
Weight
Instrument cooling method Forced air cooling. Exhaust on the left side and top panel.
DL850EV DC power supply model
(/DC option; the AC power supply specifications are the same as those shown for the standard model)
Supply format
Rated supply voltage
Permitted supply voltage 10 to 18 VDC
Power consumption
Maximum power
consumption
Standby power
(during DC standby)
Voltage input protection
circuit
Withstand voltage
External dimensions
Weight
Indicator function
Instrument cooling
method
Battery backup
Backup battery life
IM DL850E-03EN
Specifications
Ambient Temperature:
23 ± 5°C
Ambient humidity:
20 to 80% RH
Supply voltage and
Within ±1% of rating
frequency errors:
After the DL850E/DL850EV has been warmed up for 30 minutes and then calibration has been
performed
1 year
At least 30 minutes
Temperature:
−20 to 60°C
Humidity:
20 to 85% RH (no condensation)
Altitude:
3000 m or less
Temperature:
5 to 40°C
Humidity:
20 to 85% RH (when the printer is not being used)
35 to 85% RH (when the printer is used)
In both cases, no condensation
Altitude:
2000 m or less
100 to 120 VAC, 220 to 240 VAC (auto switching)
90 to 132 VAC, 198 to 264 VAC
50/60 Hz
48 to 63 Hz
Built in (not replaceable)
200 VA
1500 VAC for one minute between the power supply and case
10 MΩ or more for 500 VDC between the power supply and case
Approx. 355 mm (W) × 259 mm (H) × 180 mm (D), not including the handle and protrusions
Approx. 6.5 kg (weight of the DL850E/DL850EV only without paper and with the /M2, /HD1, /C1,
and /P4 options installed)
Auto DC/AC switching (AC preferred), isolation between DC power input terminal and the DL850EV
12 VDC
1
Approx. 90 VA (typical value
):
1
Approx. 105 VA (typical value
): When driving 8 channels (High-Speed 100 MS/s, 12-Bit Isolation
1
Approx. 135 VA (typical value
): When driving 16 channels (High-Speed 100 MS/s, 12-Bit Isolation
Approx. 150 VA
1
30 mW (typical value
, DC power consumption when power is supplied both to AC and DC power)
Overcurrent detection: Breaker (15 A)
Reverse connection protection: Breaker shutdown
Undervoltage detection: Cut off at a voltage less than approx. 9.5 V
Overvoltage detection: Cut off at a voltage greater than approx. 18 V
30 VDC for one minute between the DC power input terminal and earth terminal
Approx. 355 mm (W) × 259 mm (H) × 202 mm (D),
(DL850EV with the /DC option installed, not including the handle and protrusions)
Approx. 7.4 kg (only the DL850EV with all options (/M3/C8/C10/P4/DC options)
Approx. 9.8 kg (DL850EV (with /DC option) + eight High-speed 10 MS/s, 12-Bit Isolation Modules)
Weight increase by the DC power option Approx. 800 g
Indicates the status of the power supply to the DL850EV using two LEDs
Forced air cooling. Exhaust on the left side and top panel.
In DC mode, air is vented from the left side, top panel, and rear panel.
The settings and clock are backed up with an internal lithium battery.
Approx. five years (at an ambient temperature of 25°C)
When driving 2 channels (High-Speed 10 MS/s, 12-Bit Isolation
Module × 1)
Module × 2 and High-Speed 10 MS/s 12-Bit Isolation Module × 2)
Module × 4 and High-Speed 10 MS/s 12-Bit Isolation Module × 4)
1
2
3
4
5
6
App

Index

6-17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dl850ev

Table of Contents