Connecting the Strain Module and the Bridge Head
When Using the Strain Module (701270) and the Bridge Head (701955 or 701956)
Using the cable that came with the bridge head (701955 or 701956), connect the Strain Module
(701270) and the bridge head.
Strain Module 701270
(STRAIN_NDIS)
When Using the Strain Module (701271) and the Bridge Head (701957 or 701958)
Using the cable that came with the bridge head (701957 or 701958), connect the Strain Module
(701271) and the bridge head.
Strain Module 701271
(STRAIN_DSUB)
When Using a Bridge Head with a MIL Standard (MIL-C-26482) Connector Wiring
The connector on the Strain Module (701270) is a NDIS connector.
cable (700935) by YOKOGAWA to make a MIL-NDIS conversion and connect to the Strain Module
(701270).
* Connector recommended by JSNDI (Japanese Society for Non-Destructive Inspection)
Strain Module 701270
(STRAIN_NDIS)
NDIS connector pinout
F
F
G
G
E
E
D
D
IM DL850E-03EN
Cable
Cable
Connector adapter cable
(700935) 1.5 m long
A
A: Bridge+ (positive bridge
voltage)
B
B
B: Input– (negative
C
C
measurement signal)
C: Bridge– (negative bridge
voltage)
D: Input+ (positive
measurement signal)
E: Floating common
F: Sense+ (positive bridge
voltage sensing)
G: Sense– (negative bridge
voltage sensing)
3.9 Connecting Bridgeheads
Bridgehead
701955 or 701956
Bridgehead
701957 or 701958
*
Use a connector adapter
Bridgehead with MIL standard
MIL standard connector pinout
A
A: Signal+
B
B
F
F
B: Signal–
C
C
E
E
C: Excitation+
D
D
D: Excitation–
E: NC (no connection)
F: NC (no connection)
Index
connector wiring
Corresponding
terminals
NDIS
MIL
A
C
B
B
C
D
D
A
3-37
1
2
3
4
5
6
App