YOKOGAWA DL850E User Manual page 131

Scopecorder
Hide thumbs Also See for DL850E:
Table of Contents

Advertisement

GO/NO-GO Determination I/O
You can apply an external signal to the DL850E/DL850EV's GO/NO-GO I/O terminal and perform GO/
NO-GO determination, and you can output the results of GO/NO-GO determination from the GO/NO-
GO I/O terminal.
GO/NO-GO I/O Connector
Type
The connector uses an RJ-11 modular jack. Use the external I/O cable accessory (720911; sold
separately). If you are using a commercially sold cable (four-conductor modular telephone cable),
wire the pins according to the above figure.
I/O Level
Within 0 to 5 V, threshold level: TTL
Pinout
EXT I/O
6
1
Connector on
the DL850E/DL850EV
Input Signal
START IN (Negative logic)
Use this signal to perform GO/NO-GO determination by synchronizing to an external input signal.
The signal is only valid when on the GO/NO-GO menu, Remote is set to ON. If Remote is set to
OFF, GO/NO-GO determination is performed regardless of the external signal input (the GO/ NO-
GO determination result is output).
Signal Input Circuit
START IN
120 Ω
IM DL850E-03EN
5.5 GO/NO-GO Determination I/O and External Start/Stop Input (EXT I/O)
Pin no.
Signal
1
EXT EVENT IN
2
START IN
3
GO OUT
4
NOGO OUT
5
GND
6
NC (no connection)
5 V
Equivalent to
4.7 kΩ
LCX14
100 pF
IN
Manual event. Event input occurs on low edge.
IN
Starts on low edge
OUT
Active low (GO)
OUT
Active low (NO-GO)
HIGH
LOW
START
1
2
3
4
5
6