Appendix 7 High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS) (701260) Specifications
Item
Specifications
Bandwidth limit
Select from Full, 10 kHz,1 kHz, and 100 Hz
Cut-off characteristics: −12 dB/OCT (Typical
Probe attenuation setting
Voltage probe: 1:1, 10:1, 100:1, 1000:1
Current probe: 1 A:1 V, 10 A:1 V (for the 701932/701933), 100 A: 1 V (for the 701930/701931)
Compatible probes/
Connection cable (for high voltage 1:1): Recommended 1
cables
Voltage probe: Recommended 2
Current probe (power can be supplied from the DL850E/DL850EV. Option)
1 Value measured under standard operating conditions.
Combined with the 700929, 702902, or 701947
700929
702902
701947
Direct input (cable that does not comply with the safety standards)
BNC
8
L
10 The typical value is a representative or standard value. It is not strictly warranted.
• To receive high voltage with the module, use a 1:1 safety cable (combination of 701901
and 701954), isolated probe (700929, 701947), or passive probe 702902.
• Note that the overvoltage tolerance of the low and high sides of this module may differ
depending on the combination of the connected probe, measurement lead, and clip.
French
• Pour recevoir une tension élevée avec le module, utiliser un câble de sécurité 1:1
(combinaison de 701901 et 701954), une sonde isolée (700929, 701947) ou une sonde
passive 702902.
• Noter que la tolérance de surtension des du bas et du haut de ce module peut être
différente suivant la combinaison de sonde reliée, de fil de mesure et de pince.
IM DL850E-03EN
701901 (isolated type BNC-safety alligator clip adapter ×2: For measuring 850 V (DC+ACpeak) or
less), 701954 (alligator clip (dolphin type) red/black 2-piece set) is required separately
700929 (10:1 safety probe), 20 to 45 pF: For measuring 1000 V (DC+ACpeak) or less
702902 (10:1 safety probe), 25 to 40 pF: For measuring 1000 V (DC+ACpeak) or less
701947 (100:1 probe), 15 to 45 pF: For measuring 1000 V (DC+ACpeak) or less
701930 (150 A), 701931 (500 A), 701932 (30 A), 701933 (30 A)
Combined with the 701901+701954
H
2
3
L
4
H
9
Withstand voltage: 3700 Vrms for 1 minute
Allowable transient surge voltage: ±5200 Vpeak
(between earth and input)
AVERTISSEMENT
10
)
701901
701954
5
L
7
WARNING
H
6
App-33
1
2
3
4
5
6
App
Index