• For the 701267 (HV (with RMS))
Maximum input voltage (at a frequency of 1 kHz or less)
• Combination of 758933 measurement leads and 701954 alligator clips
850 V (DC + ACpeak)
• Direct input (cable that does not comply with the safety standards)
42 V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• Combination of 758933 measurement leads and 701954 alligator clips
(Use caution because the overvoltage tolerance differs between the low and high sides.)
700 Vrms (CAT II) on the H side,
• Direct input (cable that does not comply with the safety standards)
42 V (DC + ACpeak) (CAT II, 30 Vrms)
For the 701261, 701262, and 701265
(Red)
Input
terminal
(Black)
For the 701267
With the 758933 and 701954
(Red)
Input
terminal
(Black)
Over-Range Indication
If over-range is indicated, the DL850E/DL850EV may be receiving a voltage higher than the
observed waveform or measured waveform values. To prevent electric shock, change the
vertical scale with the SCALE knob so that the entire amplitude of the waveform is displayed
within the waveform display area, and check the input voltage level.
French
• Toujours mettre l'appareil à mesurer hors tension avant de le brancher sur l'instrument. Il
est extrêmement dangereux de brancher un câble de mesure lorsque l'appareil à mesurer
est sous tension.
• Lors de la mesure de tension élevées, ne pas relier une borne de type enfichable, avec
des parties conductrices exposées, à la borne d'entrée à utiliser comme un fil de mesure. Il
est très dangereux que le connecteur se détache.
IM DL850E-03EN
4
400 Vrms (CAT II)
H
1
2
L
4
3
758933
701954
L
5
Over-range indication
Indicates the number of the channel that over-range is occurring on.
Channel indication when over-range is occurring on multiple channels
Indicates the smallest number among the channels that over-range is
occurring on.
AVERTISSEMENT
3.7 Connecting Measurement Leads
6
5
on the L side
7
Direct input (cable that does not comply
with the safety standards)
H
(Red)
6
Input
terminal
L
(Black)
3
5
H
7
Index
3-31
1
2
3
4
5
6
App

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dl850ev

Table of Contents