3.3 Installing Input Modules
Types of Input Modules
The following types of input modules are available.
High-Speed 10 MS/s, 12-Bit Isolation Module
High-Speed High-Resolution 1 MS/s, 16-Bit Isolation Module
High-Speed 10 MS/s, 12-Bit Non-Isolation Module
High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS)
Universal (Voltage/Temp.) Module
Universal (Voltage/Temp.) Module (with AAF)
Temperature, High Precision Voltage Isolation Module
Strain Module (NDIS)
Strain Module (DSUB, Shunt-Cal)
Acceleration/Voltage Module (with AAF)
Frequency Module
High-Speed 100 MS/s, 12-Bit Isolation Module
High-Speed 100 MS/s, 12-Bit Isolation Module
16-CH Voltage Input Module
16-CH Temperature/Voltage Input Module
Logic Input Module
CAN Bus Monitor Module
CAN & LIN Bus Monitor Module
SENT Monitor Module*
4-CH 1 MS/s, 16-Bit Isolation Module
* The CAN bus monitor, CAN & LIN bus monitor, and SENT monitor modules can be used on the DL850EV. They
cannot be used on the DL850E.
Precautions to Be Taken When Installing or Removing Input
Modules
If you replace one installed input module with another, the settings for the channel will be reset to their
defaults when the power is turned on. If you want to keep the settings, specify a save destination and
save them.
Installation Procedure of Modules
1.
Make sure that the power switch on the left side panel of the instrument is turned off.
2.
Check the channel number displayed on the input module installation slot on the right side
panel of the instrument, and then install the input module along the guide.
Holding the handles on the top and bottom of the input module, press hard until it clicks in place.
If there is a cover panel on the slot in which to install the module, remove the cover panel, first.
3.
Firmly fasten the screws that came with the instrument in two places: the top and bottom of the
input module.
Screw tightening torque: 0.6 N•m
4.
Turn the instrument's power switch on.
5.
In the overview screen, check that the name of the module that you installed is displayed
correctly at the appropriate slot. If it is not correct, remove the module according to the steps
in "Removal" shown below, and reinstall the module according to steps 1 to 3 shown above. To
display the overview screen, see section 19.4,
user's manual.
3-8
*
*
Guide
701250 (HS10M12)
701251 (HS1M16)
701255 (NONISO_10M12)
701267 (HV (with RMS))
701261 (UNIVERSAL)
701262 (UNIVERSAL (AAF))
701265 (TEMP/HPV)
701270 (STRAIN_NDIS)
701271 (STRAIN_DSUB)
701275 (ACCL/VOLT)
701281 (FREQ)
720210 (HS100M12)
720211 (HS100M12)
720220 (16CH VOLT)
720221 (16CH TEMP/VOLT)
720230 (LOGIC)
720240 (CAN MONITOR)
720241 (CAN & LIN)
720243 (SENT)
720254 (4CH 1M16)
Viewing System Information (Overview)," in the
"