Connecting Differential Probes
When using Yokogawa differential probes (700924, 700925, 701926), connect the BNC output
connector to the oscilloscope's input terminal. Also, be sure to connect the GND lead to the functional
ground terminal of the DL850E/DL850EV. If necessary, use the auxiliary grounding lead extension.
For details on the connection procedure, see the manual that came with the differential probe.
In Using the High Voltage Differential Probes
Be sure to connect the ground lead of the differential probe (700924, 700925, 701926) to
the functional grounding terminal on the right side panel of the DL850E/DL850EV before
connecting to the measurement target. Not doing so may cause high voltage to appear in the
BNC connector of the differential probe.
French
Utilisation de sondes différentielles haute tension
Bien relier le fil de terre de la sonde différentielle (700924, 700925, 701926) à la borne de
mise à la terre opérationnelle, sur le panneau droit de DL850E/DL850EV avant de le relier à
la cible de la mesure. Le fait de ne pas respecter cette consigne risque d'entraîner l'apparition
d'une tension élevée au niveau du connecteur BNC de la sonde différentielle.
IM DL850E-03EN
Functional ground terminal
GND lead
WARNING
AVERTISSEMENT