Item
Table of Cutoff Frequency
Characteristics of the Anti-
Aliasing Filter (AAF)
1 Value measured under standard operating conditions.
Combined with the 10:1 passive probe
(701940)
2
4 The typical value is a representative or standard value. It is not strictly warranted.
IM DL850E-03EN
Specifications
When the filter is set to Auto, the anti-aliasing filter and low-pass filter are automatically set
according to the sample rate.
Sample Rate
AAF
100 kS/s
40 kHz
50 kS/s
20 kHz
20 kS/s
8 kHz
10 kS/s
4 kHz
5 kS/s
2 kHz
2 kS/s
800 Hz
1 kS/s
400 Hz
500 S/s
200 Hz
200 S/s
80 Hz
100 S/s
40 Hz
50 S/s
20 Hz
20 S/s to 5 S/s
20 Hz
2 S/s or less
20 Hz
Ext sample
40 kHz
Direct input (cable that does not comply
with the safety standards)
H
BNC
L
3
6.13 Module Specifications
Low-pass filter
OFF
OFF
OFF
4 kHz
4 kHz
4 kHz
400 Hz
400 Hz
400 Hz
40 Hz
40 Hz
40 Hz
40 Hz
OFF
H
2
L
3
1
2
3
4
5
6
App
Index
6-37