5.6

IRIG Signal Input (IRIG option)

Only apply signals that meet the following specifications. Signals that do not meet the
specifications may damage the DL850E/DL850EV, because of factors such as excessive
voltage.
French
Appliquer uniquement des signaux respectant les spécifications ci-dessous. Les signaux
ne respectant pas les spécifications risquent d'endommager DL850E/DL850EV, à cause de
facteurs tels que la surtension.
IRIG Signal Input Terminal
IRIG
You can use an IRIG (Inter Range Instrumentation Group) signal to set the time on the DL850E/DL850EV.
Item
Input connector
Number of input connectors
Supported IRIG signals
Input impedance
Maximum input voltage
Used for
Clock sync range
Post-sync accuracy
5-10
CAUTION
ATTENTION
Specifications
BNC
1
A002, B002, A132, and B122
You can switch between 50 Ω and 5 kΩ.
±8 V
Synchronizing the DL850E/DL850EV time
Synchronizing the sample clock
±80 ppm
No drift from the input signal