Sent Monitor Module (720243) - YOKOGAWA DL850E User Manual

Scopecorder
Hide thumbs Also See for DL850E:
Table of Contents

Advertisement

6.13 Module Specifications

SENT Monitor Module (720243)

Item
Standard operating
condition
Number of input ports
Maximum data update rate
Input format
Input connector
Input impedance
Supported protocol
Supported clock tick span
Supported number of nibbles 1 to 8
Number of analysis sub
channels
Low level input voltage
High level input voltage
Maximum input voltage
Maximum allowable common
mode voltage
Withstand voltage
Allowable transient surge
voltage (instantaneous)
Insulation resistance
Input status indication
function
Probe attenuation setting
Compatible probes/cables
1
Typical value represents a typical or average value. It is not strictly warranted.
Combined with the 700929 or 702902
700929
702902
Withstand voltage: 1500 Vrms for 1 minute
Allowable transient surge voltage (between earth and input): ±2100 Vpeak
6-50
pecifications
Ambient temperature: 23±5°
Ambient humidity: 20 to 80% RH
2
100 kS/s (10 µs)
Isolated unbalanced
BNC connector (isolated type)
1 MΩ±1%, approx. 35 pF
SAE J2716
1 µs to 100 µs
Resolution: 0.01 µs
Fast channel
Slow channel
Status & communication
Error trigger
Error count
1
1.5 V (typical value
)
1
3.5 V (typical value
)
Combined with the 700929 (10:1) or 702902 (10:1):
420 V (DC+ACpeak)
Combined with the (701901 + 701954) (1:1) or direct input (cable that does not comply with the
4
safety standards):
42 V (DC+ACpeak)
Working voltage of safety standard
Combined with the 700929(10:1) or 702902 (10:1):
300 Vrms (CAT II)
Combined with the (701901 + 701954) (1:1) or direct input (cable that does not comply with the
5
safety standards):
42 V (DC+ACpeak) (CAT II, 30 Vrms)
1500 Vrms for 1 minute (across each terminal and earth) (60 Hz)
±2100 Vpeak (across each input terminal and earth)
500 VDC, 10 MΩ or more (across each input terminal and earth)
Input status indication through the LEDs of each port function
When in operation:
Illuminates in green when pulse input is detected
When overdriven:
Illuminates in red when the input voltage exceeds 20 V
10:1, 1:1
Voltage probe: Recommended
700929 (10:1 safety probe), 20 to 45 pF
702902 (10:1 safety probe), 25 to 40 pF
Connection cable (for high voltage 1:1)
701901 (isolated type BNC-safety alligator clip adapter ×2), 701954 (alligator clip (dolphin type)
red/black 2-piece set) is required separately
Connection cable (for low voltage 1:1)
366926 (non-isolated type BNC-alligator clip ×2)
Combined with the (701901 + 701954) or
direct input (cable that does not comply with the safety standards)
H
BNC
2
L
3
3 ch
5 ch
1 ch (4 bit)
1 ch (5 bit)
1 ch
2
3
H
4
L
5
IM DL850E-03EN

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dl850ev

Table of Contents