YOKOGAWA DL850E User Manual page 153

Scopecorder
Hide thumbs Also See for DL850E:
Table of Contents

Advertisement

6.12 General Specifications
Item
Safety standard
Emissions
6-18
Specifications
Compliant standards
2
EN61010-1, EN61010-2-030, EN61010-031, EN 60825-1
• Overvoltage category (installation category) II
• Measurement Category: See the specifications of each module.
5
• Pollution degree 2
Approved (DL850E, DL850EV, 701250, 701251, 701255, 701267, 701261, 701262, 701265,
701270, 701271, 701275, 701281, 720210, 720211, 720220, 720221, 720230, 720240, 720241,
720243, 720254, 700986, 700987, 701953, 701955, 701956, 701957, 701958, 702911, 702912)
• Bridgehead for the strain module
Use the 701955 or 701956 with the 701270, and use the 701957 or 701958 with the 701271.
Compliant standards
EN61326-1 Class A, EN61326-2-1, EN 55011 Class A, Group 1, EMC Regulatory Arrangement
in Australia and New Zealand EN 55011 Class A, Group 1, Korea Electromagnetic Conformity
Standard ( 한국 전자파적합성기준 )
Approved (DL850E, DL850EV, 701250, 701251, 701255, 701267, 701261, 701262, 701265,
701270, 701271, 701275, 701281, 720210, 720211, 720220, 720221, 720230, 720240, 720241,
720243, 720254, 700986, 700987, 701953, 701955, 701956, 701957, 701958, 702911, 702912)
EN61000-3-2, EN61000-3-3
This product is a Class A (for industrial environments) product. Operation of this product in a
residential area may cause radio interference in which case the user is required to correct the
interference.
Test items
1. Power supply terminal noise
2. Radiation emission
3. Power supply harmonic regulation
4. Power supply voltage fluctuation and flicker
Cable conditions (DL850E/DL850EV)
Current probe
When connecting a current probe to the input terminal and probe power terminal of a module,
attach a single ferrite core
DL850EV.
GP-IB cable
Use a shielded cable that is 3 m or less in length.
USB cable
Use a shielded cable that is 3 m or less in length, and attach a ferrite core
cable closest to the DL850E/DL850EV.
Ethernet cable
Use a shielded cable that is 30 m or less in length, and attach a ferrite core
cable closest to the DL850E/DL850EV.
External clock input, external trigger input, external trigger output
Use a shielded cable that is 3 m or less in length, and attach a ferrite core
cable closest to the DL850E/DL850EV.
IRIG cable
Use a shielded cable that is 3 m or less in length, and attach a ferrite core
cable closest to the DL850E/DL850EV.
External HDD cable
Use a shielded cable that is 3 m or less in length.
Video signal output cable
Use a shielded cable that is 3 m or less in length, and attach a ferrite core
cable closest to the DL850E/DL850EV.
External I/O cable
Use a shielded cable that is 3 m or less in length, and attach a ferrite core
cable closest to the DL850E/DL850EV.
DC power cable
7
Attach a ferrite core
to the DC power cable.
Probe power cable
Use the dedicated cable, and attach a ferrite-core
DL850E/DL850EV by passing the cable twice through the core.
Example of passing the cable through twice
Cable conditions (input module)
50 Ω terminator 700976 (for the 701281, 720211, 720243, and 720254)
Isolated probe 700929 (for the 701250, 701251, 701255, 701275, and 720210)
6
Attach a ferrite-core
cable twice through the core.
Twisted pair cable for the 701261, 701262, and 701265
Use a cable that is 3 m or less in length, and attach a ferrite-core
closest to the DL850E/DL850EV by passing the cable twice through the core.
3
6
to both cables on the side of the cables closest to the DL850E/
to the side of the cable closest to the DL850E/DL850EV by passing the
4
6
to the side of the
6
6
to the side of the
6
to the side of the
7
to the side of the
6
to the side of the
6
to the side of the cable closest to the
6
to the side of the cable
to the side of the
IM DL850E-03EN

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dl850ev

Table of Contents