Item
Input voltage range (±FS)
Input impedance
Input coupling settings
Probe attenuation setting
Minimum voltage width for pulse
detection
Bandwidth limit
Threshold
Hysteresis
Preset function
Slope selection
Lower −3 dB point when AC
coupled
Chatter elimination function
Input status indication function
Compatible probes/cables
1 Value measured under standard operating conditions.
Combined with the 700929, 702902, or 701947
700929
702902
701947
Withstand voltage: 1500 Vrms for 1 minute
Allowable transient surge voltage (between earth and input): ±2100 Vpeak
6 Input waveform of 1 Vpp, rectangular wave, rise/fall time within 1 ms (input range: ±10 V, bandwidth limit: Full, and hysteresis: ±1%)
7 Input waveform of 90 Vrms, sine wave (input range: AC100 V, bandwidth limit 100 kHz, and hysteresis: ±1%)
8 Input waveform of 1 Vpp, rectangular wave, rise/fall time within 5 ns (input range: ±10 V, bandwidth limit: Full, and hysteresis: ±1%)
9 Typical value represents a typical or average value. It is not strictly warranted.
IM DL850E-03EN
Specifications
• When in pulse width measurement mode
Measurement accuracy is specified according to the measurement range and input pulse width
[Definition of measurement accuracy]
±(0.05% of 10 div + accuracy dependent on the input pulse width)
[Accuracy dependent on the input pulse width]
Input pulse width of 500 µs or greater:
Input pulse width of 20 µs to 500 µs:
Input pulse width of 10 µs to 20 µs:
Input pulse width of 10 µs or less:
When using 1:1 probe attenuation:
1 MΩ ± 1 approx. 35 pF
Pull-up function:
10 kΩ, approx. 5 V (pull-up can be turned ON only when the input
is set to Pull-up 5 V)
AC, DC
10:1, 1:1
200 mV
P-P
Select from Full, 100 kHz, 10 kHz, 1 kHz, and 100 Hz
Cutoff characteristics:
−12 dB/OCT (typical value
Set within the FS of the voltage range. Set in units of 1% of the FS.
Select ±1%, ±2.5%, or ±5% of the FS of the voltage range
Logic (5 V/3 V/12 V/24 V), electromagnetic pickup, zero crossing, pull-up, AC100 V, AC200 V,
and user-defined
Select rising or falling
0.5 Hz or less (0.05 Hz or less when using the 700929 or 702902, 0.005 Hz or less when
using the 701947) (typical value
OFF or 1 to 1000 ms (1 ms resolution)
Eliminates the chatter that occurs such when the contact input is turned ON/OFF.
Can discard the signal changes over the specified interval.
Input status indication through the LEDs of each channel function
When in operation:
Illuminates in green when pulse input is detected
When overdriven:
Illuminates in red when the input voltage exceeds the range
Connection cable (1:1): Recommended 1
366926
Voltage probe: Recommended 2
700929 (10:1 safety probe), 20 to 45 pF
702902 (10:1 safety probe), 25 to 40 pF
701947 (100:1 probe), 15 to 45 pF
Combined with the (701901 + 701954) or
direct input (cable that does not comply with the safety standards)
H
BNC
2
3
L
6.13 Module Specifications
8
0.05% of the input pulse width
0.1% of the input pulse width + 0.1 µs
0.2% of the input pulse width + 0.1 µs
0.5% of the input pulse width + 0.1 µs
±1 V, ±2 V, ±5 V, ±10 V, ±20 V, ±50 V (±FS)
9
)
9
)
H
4
5
L
1
2
3
4
5
6
App