Note
You can install up to four 720210 modules, and these modules must be installed in the top slots.
You can install up to four 720234 SENT monitor modules (SENT), and these modules must be installed in
the bottom slots.
You can install a total of two 720240 CAN bus monitor modules (CAN MONITOR) and 720241 CAN & LIN
bus monitor modules (CAN & LIN). These monitor modules can only be installed in slots 7 and 8.
Removal
1.
Make sure that the instrument's power switch is off.
2.
Loosen the two screws that are fastened to the input module.
3.
Hold the two handles at the top and bottom of the input module, and pull it out.
About the High-speed, 100 MS/s, 12-bit Isolated Module
Safety Precautions for Laser Products
The High-Speed 100 MS/s, 12-Bit Isolation Module (720210 (HS100M12) and 720211 (HS100M12))
uses an internal laser light source. The 720210 (HS100M12) and 720211 (HS100M12) are a class 1
laser product as defined by IEC60825-1: Safety of Laser Products—Part1: Equipment Classification.
In addition, this instrument complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant
to Laser Notice No. 50, dated June 24, 2007.
High-Speed 100 MS/s, 12-Bit Isolation Module (720210 (HS100M12) and 720211
(HS100M12))
The following information is printed on the side.
DL850E/DL850EV
The following information is printed on the rear.
IM DL850E-03EN