YOKOGAWA DL850E User Manual page 68

Scopecorder
Hide thumbs Also See for DL850E:
Table of Contents

Advertisement

• For the 701255 (NONISO_10M12)
This module is non-isolated. Be sure to fasten the module screws when measuring a
voltage above 42 V on this module. In addition, use the 701940 (10:1) dedicated non-
isolated passive probe.
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the 701940 (10:1) passive probe
600 V (DC + ACpeak)
• Direct input (cable that does not comply with the safety standards)
200 V (DC + ACpeak) (as a value that meets the safety standard)
250 V (DC + ACpeak) ( maximum allowable voltage, as a value that does not damage
• For the 701275 (ACCL/VOLT)
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the 701940 (10:1) passive probe
comply with the safety standards)
42 V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• Combined with the 701940 (10:1) passive probe
comply with the safety standards)
42 V (DC + ACpeak) (CAT II, 30 Vrms)
• For the 701281 (FREQ)
Maximum input voltage (at a frequency of 1 kHz or less)
• In combination with an isolated probe (10:1) 700929 or (100:1) 701947, or a passive
probe (10:1) 702902
420 V (DC + ACpeak)
• Safety cable (1:1) (combined with 701901+701954)
comply with the safety standards)
42 V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• In combination with an isolated probe (10:1) 700929 or (100:1) 701947, or a passive
probe (10:1) 702902
300 Vrms (CAT II)
• Safety cable (1:1) (combined with 701901+701954)
comply with the safety standards)
42 V (DC + ACpeak) (CAT II, 30 Vrms)
• For the 720243 (SENT)
Maximum input voltage (at a frequency of 1 kHz or less)
• In combination with an isolated probe (10:1) 700929 or a passive probe (10:1) 702902
420 V (DC + ACpeak)
• Safety cable (1:1) (combined with 701901+701954)
comply with the safety standards)
42 V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• In combination with an isolated probe (10:1) 700929 or a passive probe (10:1) 702902
300 Vrms (CAT II)
• Safety cable (1:1) (combined with 701901+701954)
comply with the safety standards)
42 V (DC + ACpeak) (CAT II, 30 Vrms)
IM DL850E-03EN
7
the instrument when applied)
7
5
8
6
1
5
2
6
5
6
3.5 Connecting Probes
5
or direct input (cable that does not
or direct input (cable that does not
3
or direct input (cable that does not
4
or direct input (cable that does not
3
or direct input (cable that does not
4
or direct input (cable that does not
1
2
3
4
5
6
App
Index
1
2
3-19