13.6 General Specifications
Construction
Item
Mounting
Mounting angle
Allowable panel thickness
Material
Color
Front panel
External dimensions
Weight
Normal Operating Conditions
Item
Supply voltage
Power supply frequency
Ambient temperature
Ambient humidity
Vibration
Shock
Magnetic field
Noise
DC voltage
Thermocouple
RTD
Common mode noise
Maximum noise voltage between channels
Mounting position
Warm-up time
Installation location
Operating altitude
Power Supply
Item
Rated supply voltage
Allowable power supply voltage range
Rated power supply frequency
Power consumption
Allowable interruption time
Isolation
Item
Specifications
Insulation resistance Between the Ethernet, RS-422/485, and insulation terminals and earth: 20 MΩ or greater at 500 VDC
Withstand voltage
Between the power terminal and earth:
Between the contact output terminal and earth:
Between the measurement input terminal and earth: 1500 VAC at 50/60 Hz for one minute
Mutually between measurement input terminals:
Between the remote input terminal and earth:
Between the pulse input terminal and earth:
Ground
Grounding resistance: 100 Ω or less
IM 04L42B01-01E
Specifications
Flush panel mounting (on a vertical plane) (excluding the desktop type)
Inclined backward up to 30 degrees from a horizontal plane.
2 to 26 mm
Case:
Metal plate
Bezel and display cover:
Polycarbonate
Case:
Grayish blue green (Munsell 2.0B5.0/1.7 or equivalent)
Bezel:
Charcoal gray light (Munsell 10B3.6/0.3 or equivalent)
Water and dust proof: Complies with IEC529-IP65 and NEMA No.250 TYPE 4 (except External
Icing Test) (Style number 2 or later), except for side-by-side mounting
288(W) × 288(H) × 226(D) mm (D: depth from the panel mounting plane)
DX2004 and DX2010: Approx. 6 kg. DX2008 and DX2020: Approx. 6.3 kg. DX2030: Approx. 6.9
kg, DX2040 and DX2048: Approx. 7.3 kg excluding options
Specifications
90 to 132, 180 to 250 VAC
50 Hz ± 2%, 60 Hz ± 2%
0 to 50°C
20 to 80%RH (at 5 to 40°C), 10 to 50% (at 40 to 50°C)
2
10 to 60 Hz, 0.2 m/s
Not allowed
400 A/m or less (DC and 50/60 Hz)
Normal mode (50/60 Hz)
The peak value including the signal must be less than 1.2 times the measuring range.
The peak value including the signal must be less than 1.2 times the measuring thermal electromotive force.
50 mV or less
250 VACrms or less for all ranges (50/60 Hz)
250 VACrms (50/60 Hz) or less
Can be inclined up to 30 degrees backward. Left and right horizontal.
At least 30 minutes after power on
Indoors
2000 m or less
Specifications
100 to 240 VAC
90 to 132, 180 to 264 VAC
50 Hz to 60 Hz
Supply voltage
LCD backlight off
100 VAC
28 VA
240 VAC
38 VA
Less than 1 cycle of the power supply frequency
Normal
42 VA
54 VA
2300 VAC at 50/60 Hz for one minute
1600 VAC at 50/60 Hz for one minute
1000 VAC (50/60 Hz) for one minute (excluding the
RTD input terminal of DX2010, DX2020, DX2030,
DX2040, and DX2048)
1000 VDC for one minute
1000 VDC for one minute
1
2
3
4
5
6
7
8
9
10
11
Maximum
74 VA
100 VA
12
13
App
Index
13-19