Exchange The Inlet Valve - Agilent Technologies Infinity II SFC Binary 1260 User Manual

Table of Contents

Advertisement

Exchange the Inlet Valve

When
If internally leaking (backflow)
Tools required
Description
Wrench, 14 mm
Parts required
p/n
G1312-60066
G4302-60066
Preparations
Switch off the pump at the main power switch
1
Disconnect the solvent inlet tube at the inlet valve
(beware of leaking solvents).
3
Screw the new valve into the pump head and tighten
using the 14 mm wrench. Do not overtighten.
1260 SFC Binary Pump
Description
Passive inlet valve
Inlet Valve SFC
2
Using a 14 mm wrench, loosen the active inlet valve and
remove the valve from the pump head.
4
Reconnect the inlet tube to the valve.
Maintenance

Exchange the Inlet Valve

8
119