Replace the O-Ring on the Purge Valve

When
If the original o-ring is damaged and needs to be replaced
Tools required
p/n
8710-0510
8710-1924
OR
Parts required
#
1
1
1
Preparations
• Switch off pump at the main power switch.
• Open the doors of the module.
• Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
• Remove the purge valve from the pump head.
1
Disassemble the purge valve.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Description
Open-end wrench 1/4 — 5/16 inch
Open-end wrench 14 mm
Pair of tweezers
Toothpick
p/n
Description
5067-6595
1260 PV O-ring FKM 5/pack
01018-22707
PTFE frits (pack of 5) (OPTIONAL)
5067-4728
Seal cap (OPTIONAL)
leakages.

Replace the O-Ring on the Purge Valve

2
Remove the old o-ring from the purge valve.
3
Clean the purge valve parts.