Purge Valve; Seal Wash - Agilent Technologies 1200 Series User Manual

Preparative pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

7 Connect the long capillary (300 mm, G1361-67306) to the electromagnetic
8 Connect the medium length capillary (140 mm, G1361-67304) to the
9 Install the mixer (79835-87330) in the clips on the front panel of the bottom
10 Connect the capillary (G1361-67305) from the outlet of the female T-piece
11 Install the outlet capillary (G1361-67302) to the mixer and make the
12 Install the seal wash tubings, as shown in
13 Install the front cover.
Agilent 1200 Preparative Pump User Manual
Flow Connections for a Dual (Binary Gradient) Preparative Pump
purge valve (EMPV) of the upper pump and connect the female T-piece to
the other end of the capillary.
electromagnetic purge valve (EMPV) of the lower pump and connect the
other end of the capillary to the female T-piece.
pump.
to the mixer.
necessary connections the other side of the capillary (normally a capillary
to an injector or injection valve).
for each of the pumps - a tubing from your seal wash solvent bottle to the
seal wash pump, one from there to the pump head and one from the pump
head to waste.
Installing the Pump
Figure 13
on page 48. You need -