Replacing the Outlet Filter

Replacing the Outlet Filter

When
For removing blockages and leaks in the high pressure filter assembly. The outlet filter should be
replaced as required depending on the system usage.
Tools required
p/n
8710-0510
8710-1924
Parts required
p/n
G4204-60004
1
Remove the capillary from the outlet filter to the
pressure sensor.
198
Description
Open-end wrench 1/4 — 5/16 inch
Open-end wrench 14 mm
Torque wrench
Torque wrench head, 14 mm for torque wrench
Description
Outlet filter Quaternary Pump/Flexible Pump
2
Remove the outlet filter using a 14 mm wrench.
1290 Infinity Quaternary Pump User Manual