Auxiliary Output

Pin Out:
* Maximum load of 50 k Ω necessary to maintain scale factor.
** Internal pacemaker pulses are not represented in the auxiliary output.
2042480-001 C
Connections: Auxiliary Output
The auxiliary output is a nonisolated signal input/output meant only for
connection to another medical device that is compliant with IEC 60101-1.
Connection to a noncompliant device could result in a safety risk by exceeding
current limits.
The auxiliary output connector provides an analog ECG waveform output. The
analog ECG waveform signal is a high level (1 V/mV) representation of the
primary lead. The connector is a two-conductor 3.5 mm stereo jack that mates
with a 3.5 mm stereo plug (Shogyo SPY 1011, Shogyo SPY 1012, or equivalent).
Cable construction should be shielded, three conductor with PVC insulation or
equivalent insulating material.
Tip
Ring
Sleeve
Dynamic range
Scale Factor*
Output Impedance/Short Circuit
Protection
Frequency Response
Delay**
Dash 2500 Patient Monitor
No Connection
ECG Analog Output
Signal Return
+ 4.75V minimum
1 V/mV ± 5%
<100 Ω
0.05 Hz - 100 Hz +1/-6 dB
22 + 5 ms
A-11