Description

Action
admit (no data in
trends)
admit (data in trends),
Yes to Popup menu
asking if you want to
save trends
admit (data in trends),
No to Popup menu
asking if you want to
save trends
discharge
NOTE: If a patient is admitted while an NIBP or TEMP determination is in progress, the determination is canceled. Three different default tables
(monitoring configurations) can be set in the password-protected configuration mode. Each default table should contain settings
appropriate for a particular patient type (e.g., adult, neonate, etc.). In monitoring mode, you can choose the appropriate table for your
patient through the choose patient settings option.
2042480-001 C
Admit Patient: Description
Whenever you turn on the Monitor or return from standby, a popup window
appears with the message, CAUTION! Alarms are temporarily disabled while
patient connections are re-established. Is this a new patient you would like
to admit? Select yes or no. If there is no user interaction within 30 seconds, the
Monitor automatically defaults to no. However, if you decide to admit a patient,
the Monitor immediately resets all settings and limits based upon the
preselected configuration and turns all parameters off-line. All parameters are
auto-switched on once a patient is admitted and patient connections are made.
If you want to admit a patient during monitoring mode, use the admit option.
This performs all the same operations as selecting yes from the admit popup
window, except parameters are not turned off-line.
NOTE: If SpO
parameter does not auto-switch on.
The following table guides you through the admit/discharge procedures and
how they effect the Monitor.
Admit/Discharge Procedures
Patient
Entry in History
Information
Field
no change
PATIENT ADMITTED
no change
PATIENT ADMITTED-
no change
PATIENT ADMITTED-
DISCARD
patient's name is
PATIENT
replaced with
DISCHARGED
NO PATIENT
ADMITTED
Dash 2500 Patient Monitor
was previously off-line and a new patient is admitted, the SpO
2
Reverts to
Default
Settings?
yes
yes
SAVE
yes
yes
Entry in Trends
blank trend moves to
Previous Patient, start
trending Current Patient,
admit/discharge marker
marks current time
continue trending in Current
Patient, admit/discharge
marker at last admit/
discharge
current trend data copied to
Previous Patient, new data in
Current Patient, admit/
discharge marker marks
current time
current trend data copied to
Previous Patient, new data in
Current Patient, admit/
discharge marker marks
current time
2
5-3