2042480-001 C
MASIMO SET® SpO2: Indications and Contraindications
NOTES
A patient's vital signs may vary dramatically during the use of
cardiovascular agents such as those that raise or lower blood pressure or
those that increase or decrease pulse rate.
The Dash 2500 Patient Monitor that is labeled with MASIMO SET ®
Technology is compatible only with MASIMO SET ® sensors.
Software development, software validation, and Risk and Hazard Analysis
has been performed to a registered quality system.
User or patient-applied parts are latex-free.
Procedures
1.
Select a sensor that is appropriate for the patient and the clinical situation.
Verify that the Monitor's patient type and sensor type match. Sensor sizing
must be correct to for the SpO
WARNING
Do not use a damaged sensor or one with exposed electrical
contacts.
NOTE: To assure optimal performance, use only MASIMO SET
which are available from GE Medical Systems - Accessories and
Supplies.
2.
Connect the SpO
SpO
cable to the SpO
2
fit only the SpO
3.
Following the directions for use supplied with the sensor, apply the sensor to
the patient.
WARNINGS
Patient safety:
If you fail to apply the sensor properly, the patient's skin could
be injured or the ability of the SpO
oxygen saturation could be compromised. For example, a clip-
on sensor should never be taped shut. Taping the sensor could
damage the patient's skin or impair the venous return, thus
causing venous pulsation and inaccurate measurement of
oxygen saturation.
Excessive pressure from the sensor may cause necrosis of the
skin.
Parameter performance:
When an SpO
cuff, the SpO
SpO
readings are required during the entire blood pressure
2
determination, attach the SpO
one with the blood pressure cuff.
Dash 2500 Patient Monitor
algorithm to function properly.
2
sensor to the SpO
extender cable. Then, connect the
2
2
port on the Monitor. (The SpO
2
port.)
2
parameter to measure
2
sensor is on a limb that has a blood pressure
2
data will not be valid when the cuff is inflated. If
2
sensor to the limb opposite the
2
®
sensors,
cable is designed to
2
13-7