Alarms

Name of Alarms
Systolic High
Systolic Low
Mean High
Mean Low
Diastolic High
Diastolic Low
NIBP No
Determination*
NIBP Level Timeout*
NIBP Total Timeout*
NIBP Pump
Timeout*
NIBP Overpressure*
* Alarm message appears in SelectBox.
2042480-001 C
NIBP: Alarms
NIBP Alarms
Probable Cause
value is greater than "hi" alarm
limit
value is greater than "lo" alarm
limit
value is greater than "hi" alarm
limit
value is greater than "lo" alarm
limit
value is greater than "hi" alarm
limit
value is greater than "lo" alarm
limit
insufficient signal
remained at one cuff pressure
level for more than 1 minute
due to motion
determination time exceeded 2
minutes due to motion artifact
time for inflation was exceeded
or air leak detected
appropriate level or pressure in
the cuff has been exceeded
Dash 2500 Patient Monitor
Response
message appears in alarm message field;
parameter label, violated limit (if limits set to
display), and value in violation flash in vital
signs field; NIBP OUT of LIMITS message
appears in SelectBox
message appears in alarm message field;
parameter label, violated limit (if limits set to
display), and value in violation flash in vital
signs field; NIBP OUT of LIMITS message
appears in SelectBox
message appears in alarm message field;
parameter label, violated limit (if limits set to
display), and value in violation flash in vital
signs field; NIBP OUT of LIMITS message
appears in SelectBox
message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes
message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes
message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes
message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes
message appears in alarm message field;
parameter label flashes in vital signs field;
alarm forces values to dashes