12-16
NELLCOR® OXIMAX® SpO2: Troubleshooting
PROBLEM: The SpO
value and/or PR value is out of limits but a limit alarm is not
2
audible.
CAUSE:
The NELLCOR ® SpO
the NELLCOR ® SatSeconds
SOLUTION:
Check the patient.
Check the configuration setting for SatSeconds
the limit alarm will not be generated immediately after the limit is
violated. The length of time before the alarm is generated depends both
upon how far it is out of limits and for long it has been out of limits.
Dash 2500 Patient Monitor
parameter has been set up to delay alarms using
2
™
alarm management software.
™
. If greater than zero,
2042480-001 C