Ecg; Resp; Spo - GE Dash 2500 Operator's Manual

Patient monitor
Hide thumbs Also See for Dash 2500:
Table of Contents

Advertisement

Alarm Priority
Crisis
Warning
Procedural
Message
OFF
Parameters
Heart Rate/
Pulse

ECG

NIBP

SpO

2

RESP

TEMP
2042480-001 C
Alarms: Alarm Priorities
Audible Indications
none
none
Crisis
VTACH
Resp
approaching
Dash 2500 Patient Monitor
Record on Alarm?
yes, if option chosen
yes, if option chosen
no
no
none
Default Alarm Priorities
Warning
Limit alarms
priority
Lead fail,
Replace electrodes
Limit alarms
No determination,
priority
Over-pressure,
Pump timeout,
Total timeout,
Level timeout.
Limit alarms
Lost pulse,
priority
Sensor
disconnected,
Sensor faulty,
Sensor off
Limit alarms
Lead fail,
priority
Saturation,
Artifact
Bad Probe,
Too Hot,
Disconnected
Visual
Visual Indication
Indication
Flashing Rate
Color
red
0.5 Hz
red
0.5 Hz
red
0.5 Hz
yellow
constant on
none
none
Procedural
Message
Artifact
4-5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents