Table 3-2:
INSTALLATION ROOM
OF PRODUCT OR
COMPONENT
Examination room
Technical room (See
Notice below
Control room
The chiller is able to dissipate maximum continuously power at 3000 meters in an
ambient temperature up to 20 C, for this altitude the technical room temperature
shall not exceed 20 C.
In some cases condensation occurs and water drops from outlets and pipes of the
air conditioner in the technical room.
Therefore, it is critical to install the cabinets where there is no risk of flood from the
air conditioner.
Ensure the air outlet positions in the exam room is not in area closed to monitor
suspension carriage/rail.

2.2 Equipment Heat Output (Dissipation)

2.2.1 Equipment Heat Output tables

Refer to
Table 3-3
3-3
, refer to the appropriate component Pre–Installation Manual listed in
3,
Basic Innova LC System Compatibility.
Chapter 3 Physical Requirements of Room
ALTITUDE (meters)
IN-USE
MIN
MAX
MIN
0
3000
-30
0
3000
-30
0
3000
0
. To obtain heat output information for components not specified in
Table 3-1
Chapter 2, Section 3,
Table 3-2
STORAGE /
TRANSPORT
MAX
MIN
3000
70
3000
70
3000
N/A
Direction 5314528-2-1EN, Revision 1
. To obtain relative humidity and
Table 3-1
, refer to the appropriate
Basic Innova LC System
. To obtain altitude and
Table 3-2
Chapter 2, Section 3,
ATMOSPHERIC PRESSURE (kPa)
IN-USE
STORAGE / TRANSPORT
MAX
MIN
106
70
106
50
N/A
N/A
Chapter 2, Section
, refer to the
Basic Innova
MAX
106
106
N/A