Alarms - GE Dash 2500 Operator's Manual

Patient monitor
Hide thumbs Also See for Dash 2500:
Table of Contents

Advertisement

Alarms

Name of Alarm
SpO
High
2
SpO
Low
2
SpO
PR High*
2
SpO
PR Low*
2
SpO
Sensor Disconnected sensor disconnected from
2
SpO
Lost Pulse
2
SpO
Sensor Off
2
SpO
Signal Quality
2
SpO
Sensor Faulty
2
* Not active when the View SpO2 PR? option is set to no.
13-18
MASIMO SET® SpO2: Alarms
SpO
Alarms
2
Cause
value is greater than "hi"
alarm limit
value is less than "lo" alarm
limit
value is greater than "hi"
alarm limit as set in HR/
Pulse
value is less than "lo" alarm
limit as set in HR/Pulse
module
no or very low SpO
signal,
2
signal was lost
SpO
sensor has been
2
removed from the patient
low perfusion, excessive
patient movement, too
much ambient light or
interference from another
source is degrading the
quality of the signal
SpO
sensor or cable
2
possibly defective, cable not
connected properly
Dash 2500 Patient Monitor
Response
message appears in alarm message
field; parameter label, violated limit (if
limits set to display), and value in
violation flashes in vital signs field
message appears in alarm message
field; parameter label, PR label, violated
limit, violated PR limit (if limits set to
display), and value in violation flashes
in vital signs field
message appears in waveform and
alarm message field; parameter label
flashes in vital signs field
message appears in waveform field,
alarm message field, and in SelectBox
when Enable spot check enable is set
to yes; parameter label flashes in vital
signs field
message appears in waveform field,
alarm message field, and in SelectBox
when Enable spot check is set to yes;
parameter label flashes in vital signs
field
message appears in waveform and
alarm message field; parameter label
flashes in vital signs field
message appears in waveform and
alarm message field; parameter label
flashes in vital signs field
2042480-001 C