2.2 Equipment Heat Output (Dissipation)

2.2.1 Standard System

Refer to
Table
refer to the appropriate component Pre–Installation Manual listed in
Compatibility.
56
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
NOTICE
In some cases condensation occurs and water drops from outlets and pipes of the
air conditioner in the technical room.
Therefore, it is critical to install the cabinets where there is no risk of flood from the
air conditioner.
toTable 3-1
3-3. To obtain heat output information for components not specified in
Direction 5177951-4-1EN, Revision 1
and
Table
3-2:
2 Environmental Requirements/Limitations
Table
Chapter 2, System
3-3,