print all graphs

Choose graphs to print

Display as

view older trends

view newer trends

Time to view

Move cursor

6-10
Trends: View Patient Trends Menu: Graphs
This option lets you print all graphs currently displayed on the trends screen. It is
available when the printer is not busy, trended patient data other than an admit
mark is displayed, and the Display as option is set to graphs.
This option lets you choose which graphs you wish to print. It is used to select
which graph to print when in graphs mode.
This option lets you choose the mode in which you view the trended data. Two
modes are available: numbers or graphs.
This option lets you scroll page by page through the older trends.
This option lets you scroll page by page through the newer trends.
This option lets you view data from different times over the last 24 hours. The
time displayed above the right-hand column of the 9-column group is the same
as the time shown next to the Time to view option. When you select the Time to
view option, a box appears around the time. As you turn the Trim Knob
clockwise, the time in the box moves forward and the Viewing Window in the
scroll bar moves to the right. As you turn the Trim Knob counterclockwise, the
time in the box moves backward and the Viewing Window moves to the left.
When the Trim Knob stops, the new time in the box appears above the right-
hand column of the 9-column group and the data on the screen change
accordingly. When you have found the desired data, push the Trim Knob to
confirm your choice.
This menu option lets you view the time and values associated with trended
graphical data. Once you select this option, a cursor line appears in the trended
data area. Turn the Trim Knob to move the cursor line across the trended data.
As the cursor line touches a plotted point on the graphical trends, the
corresponding time and values appear in the CURSOR column. This option is
unavailable when no trend data is available.