Parameter
HR/Pulse
50 bpm (adult)
60 bpm (ped)
RESP
6 br/min (adult)
10 br/min (ped)
NIBP
Systolic: 80 mmHg (adult)
Systolic: 70 mmHg (ped)
Diastolic: 30 mmHg (adult)
Diastolic: 30 mmHg (ped)
Mean: 40 mmHg (adult)
Mean: 40 mmHg (ped)
SpO
% Saturation: 90 (adult)
2
% Saturation: 92 (ped)
2042480-001 C
Alarms: System Fault Alarm
Default Alarm Limits

Adult/Ped

Low
150 bpm (adult)
30 br/min (adult)
Systolic: 200 mmHg (adult)
Systolic: 150 mmHg (ped)
Diastolic: 120 mmHg (adult)
Diastolic: 90 mmHg (ped)
Mean: 140 mmHg (adult)
Mean: 100 mmHg (ped)
% Saturation: 100 (adult)
% Saturation: 100 (ped)
Dash 2500 Patient Monitor
High
160 bpm (ped)
60 br/min (ped)

Neonate

Low
90 bpm
200 bpm
15 br/min
100 br/min
Systolic:
Systolic:
40 mmHg
100 mmHg
Diastolic:
Diastolic:
20 mmHg
60 mmHg
Mean:
30 mmHg
70 mmHg
% Saturation: 88
% Saturation: 100
High
Mean:
4-13