Specifications; Neonate; Adult/Ped - GE Dash 2500 Operator's Manual

Patient monitor
Hide thumbs Also See for Dash 2500:
Table of Contents

Advertisement

Specifications

ECG
Heart rate accuracy
Time to alarm
NELLCOR SpO
2
Range
Accuracy and Tolerance
Low Perfusion
MASIMO SpO
2
Range
Accuracy and Motion Tolerance
Without Motion
With Motion
Noninvasive Blood Pressure
Range

Adult/ped

Neonate

Accuracy
Alarm Limits
Range
8-6
Heart Rate/Pulse: Specifications
Specifications
30 to 300 bpm
± 3 bpm or 3% of reading, whichever is greater
high heart rate < 10 s per AAMI EC13 - 2002
low heart rate < 10 s per AAMI EC13 - 2002
cardiac standstill < 10 s per AAMI EC13 - 2002
tachycardia waveforms < 10 s per AAMI EC13 - 2002
20 - 250 bpm
20 to 250 bpm ±3 digits
20 to 250 bpm ±3 digits
25 - 240 bpm
25 to 240 bpm ±3 digits
normal physiologic range
25 to 240 bpm ±5 digits
30 - 200 bpm
30 - 220 bpm
± 3.5%
30 - 250 bpm
Dash 2500 Patient Monitor