Indications And Contraindications - GE Dash 2500 Operator's Manual

Patient monitor
Hide thumbs Also See for Dash 2500:
Table of Contents

Advertisement

Indications and Contraindications

13-4
MASIMO SET® SpO2: Indications and Contraindications
The SpO
parameter is indicated for use in continuous, noninvasive monitoring
2
of functional oxygen saturation and in providing pulse rate data as a
component of the Dash 2500 Patient Monitor. This device is not designed, sold,
or intended for use except as indicated.
WARNINGS
Do not use the SpO
imaging (MRI). Adverse reactions include potential burns to
patients as a result of contact with attachments heated by the
MRI radio frequency pulse, potential degradation of the
magnetic resonance image, and potential reduced accuracy of
SpO
measurements. Always remove oximetry devices and
2
attachments from the MRI environment before scanning a
patient.
The use of cardio-green and other intravascular dyes at certain
concentrations may affect the accuracy of the SpO
measurement.
The SpO
oxygen saturation. Significant levels of dysfunctional
hemoglobins such as carboxyhemoglobin or methemoglobin
may affect the accuracy of the SpO
A functional tester cannot be used to assess the accuracy of a
pulse oximeter probe or a pulse oximeter monitor.
Dash 2500 Patient Monitor
parameter during magnetic resonance
2
parameter is calibrated to read functional arterial
2
2
measurement.
2
2042480-001 C