Advertisement

Advanced settings

View signal strength bar?

13-10
MASIMO SET® SpO2: SpO2 Menu
NOTE: The auto-set option is not available for SpO
auto-set option to set the alarm limits for oxygen saturation. However,
when SpO
Pulse menu will set the alarm limits around the SpO
adult, pediatric and
neonatal
Heart
Rate/Pulse
Low
adult,
30 - 245 (bpm)
pediatric,
neonatal
* Only when SpO
is the source for HR/Pulse or the View SpO2 PR? option is set to Yes.
2
This option lets you view the signal strength bar, which is an indication of the
strength of the electronic signal produced by the photodiode.
Dash 2500 Patient Monitor
is the source for HR/Pulse, using auto-set in the Setup HR/
2
SpO
Alarm Limits
2
saturation:%
Low
20% - 95%
SpO
Alarm Limits: PR*
2
Adjustable
High
35 - 250 (bpm)
. You cannot choose the
2
PR value.
2
High
25% - 100%
Auto-set
choice: 20%, 30%, 40%
Low
High
current reading
current reading
minus chosen
plus chosen
percentage
percentage
2042480-001 C