Specifications - GE Dash 2500 Operator's Manual

Patient monitor
Hide thumbs Also See for Dash 2500:
Table of Contents

Advertisement

Specifications

Measurement Range
SpO
2
Pulse Rate
Perfusion Range
Accuracy and Motion Tolerance
Saturation
Without Motion - Adult/Pediatric*
Without Motion - Neonate*
With Motion - Adult/Pediatric/Neo**†
Low Perfusion‡
Pulse Rate
Without Motion
With Motion
®
* The MASIMO SET
SpO
parameter with LNOP-Adt sensors has been validated for no motion accuracy in human blood studies on healthy adult
2
volunteers in induced hypoxia studies in the range of 70-100% SpO
or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
®
**The MASIMO SET
SpO
parameter with LNOP-Adt sensors has been validated for motion accuracy in human blood studies on healthy adult
2
volunteers in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a nonrepetitive
motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO
and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the
population.
®
†The MASIMO SET
SpO
parameter with LNOP-Neo Pt sensors has been validated for neonatal motion accuracy in human blood studies on
2
neonates while moving the neonate's foot at 2 to 4 cm against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus, one
standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
®
‡The MASIMO SET
SpO
parameter has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and
2
MASIMO's simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to
100%. This variation equals plus or minus, one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
NOTE: LNCS sensors using LNCS cable technology offer nearly the same performance as the LNOP flex-circuit
design. The LNOP single patient sensor line is still the industry "gold standard", while LNCS is the best cabled
sensor alternative.
2042480-001 C
MASIMO SET® SpO2: Specifications
Specifications
1 to 100%
25 to 240 bpm
0.02 to 20%
70 to 100% ±2 digits
70 to 100% ±3 digits
70 to 100% ±3 digits
70 to 100% ±2 digits
0 to 69% unspecified
25 to 240 bpm ±3 digits
normal physiologic range
25 to 240 bpm ±5 digits
against a laboratory co-oximeter and ECG monitor. This variation equals plus
2
Dash 2500 Patient Monitor
against a laboratory co-oximeter
2
13-13