9316 Wavelength Out Of Range; 9317 No Peak: Low Hc Lamp Energy; 9318 No Peak: High Hc Lamp Energy - Agilent Technologies 55B AA User Manual

Hide thumbs Also See for 55B AA:
Table of Contents

Advertisement

Troubleshooting/Errors

9316 Wavelength out of range

The Instrument is reporting that the current method has specified a
wavelength that is out of range for the Photomultiplier Tube (PMT).
Possible causes include:
Currently available PMTs:

9317 No Peak: Low HC Lamp Energy

Low hollow cathode lamp energy.
The instrument Gain had been raised to its highest value but the
hollow cathode lamp signal was too low.
Possible causes include:

9318 No Peak: High HC Lamp Energy

High hollow cathode lamp energy.
The instrument Gain had been set to its lowest value but the hollow
cathode lamp signal was too high.
Possible causes include:
126
A method wavelength suitable for a Wide Range PMT was
specified but the instrument is currently fitted with a Short
Range PMT. Use a valid wavelength.
Wide range: up to 900 nm
Hollow cathode lamp misaligned
The light path is obstructed in the sample compartment
Faulty hollow cathode lamp
Faulty PMT or related circuits.
Hollow cathode lamp misaligned
Faulty hollow cathode lamp
Faulty PMT or related circuits.
Agilent 55B AA Spectrometer User's Guide