Pull down Data > Transforms > 1 Single-Ended, 1 Balanced from the menu. The new Test
Set Connections box opens to show the proper connections:
Click OK to close the Test Set Connections box.
Setup > 2-Port (Ports 1 and 3) / (Ports 2 and 4)
With a 4-port calibration, it is possible to make 2-port measurements. This reduces the
number of sweeps from 8 (for a 4-port device) to 2. Test ports 1 and 3, or test ports 2 and 4 are
used in a 2-port measurement. 1-port reflection measurements can be made on balanced
devices in the port 2 and 4 configuration only.
Select Setup > 2-Port (Ports 1 and 3) or Setup > 2-Port (Ports 2 and 4) to change the
configuration from 4-Port to 2-Port.
The Test-Set Connections box for ports 1 and 3, or ports 2 and 4 opens to show the proper
connections:
Click OK to close the Test Set Connections box.
You can proceed with making a 2-port measurement.