Menu Reference
Setup Menu
NOTE
Port connectors are defined by the cal kits assigned to them in the Calibration
Kit Selection window (see
Click the Use in Reverse Orientation box to change the orientation of the adapter, probe, or
fixture relative to the DUT port. This option does not work when Rotate Reference Plane is
selected. It only works with selected de-embed files.
In the screen below, the fixture has been applied in forward orientation on the left-side ports
of the DUT, and reverse orientation on the right-side ports of the DUT. Reverse application is
noted by the (R) indicator.
NOTE
If you open de-embedding files that conflict with the current de-embedding, the
software warns you:
Click the
button to check the de-embedding status. Click each DUT port to add the
rotation to that port. Click Perform De-embed Now to apply the rotation values to the
current data set. Click OK to accept the settings and close the De-embedding window.
116
"Setup > Cal Kit Selection..." on page
112).