Setup Menu
De-Embedding Status
Data files saved after de-embedding has been applied to the data are saved with the
de-embedding in place. The software makes note of the de-embedding status when you
subsequently load the data file. If the current data has had de-embedding applied (the current
data can be either measured or loaded data), the following section is added to the
De-Embedding dialog box.
Click Show De-embedding Used to see the details of the de-embedding.
Click OK to return to the De-Embedding dialog box.
Click Change De-Embed definitions to match stored data to apply the de-embedding
definitions from a stored file to the DUT for further measurements.
The software will look for the files that were used for de-embedding. The full path name and
time stamp must match exactly. If they do not match, you will be warned and must re-define
de-embedding as described above.