How to Measure Simultaneous Conjugate Match of a
Differential 2-Port Device
The software includes a set of predefined display configurations, complete with the necessary
formulas, to help in measuring the simultaneous conjugate match of a differential two-port
device. To use these configurations:
1. Either load previously saved data (see
measurement on a differential device.
2. Open the Display Configuration dialog box. See
page
121. Click LOAD ALL to open the simultaneous conjugate match display
configuration file, called Mixed Mode Parameters.mpc. This file is normally copied to
Multiport/User Config on installation.
3. Click Set Display for the Mixed Mode Parameters Config Identifier.
4. Click the User-Defined button on the main screen.
The Equations 1 and Equations 2 tabs on the user-defined display contain component
equations used in the Simultaneous Conj Match display.
Chapter 3

How to Measure Simultaneous Conjugate Match of a Differential 2-Port Device

"File > Open > Data..." on page
Using the Multiport Test System
"Setup > Display Configuration..." on
101), or make a new
71