To test offset accuracy

This test checks the vertical offset accuracy.
Specification: ±(1.00% of channel offset + 1% of full scale) at full-resolution channel scale
Equipment Required
Equipment
Critical Specifications
0.5 V to 2 Vdc, ±1 mV accuracy
Power Supply
Digital Multimeter
Better than 0.1% accuracy
(DVM)
Cables (2)
BNC
Adapters (2)
BNC (f) to banana (m)
Adapters (2)
BNC tee (m)(f)(f)
Blocking capacitor
0.18
Shorting cap
BNC

Procedure

Connect the equipment.
1
The cabling is the same as that used for the voltage measurement accuracy test. See figure 3-4.
a Use a BNC-to-banana adapter to connect a BNC cable to the power supply.
b Connect a BNC tee to the other end of the cable and connect the tee to channel 1 of the
oscilloscope.
c Connect another BNC cable to the tee at the oscilloscope and connect a BNC tee to the
other end of the cable.
d Connect the blocking capacitor to the BNC tee and connect the BNC short to the blocking
capacitor.
e Connect a BNC-to-banana adapter to the same BNC tee and connect the adapter to the
DVM input.

Press Default Setup to set the oscilloscope to default conditions.

2
Select Acquisition from the Setup menu.
3
Select Equivalent Time under Sampling Mode. Enable Averaging, and set the # Points
4
for averaging to 32. Click Close.
Setup is the same as that for the voltage measurement accuracy test. See figure 3-5.
Use the following table for steps 6 through 12.
5
Volts/div
Position
200 mV
2.00000 V
100 mV
1.00000 V
50 mV
500.000 mV
Select Channel 1 from the Setup menu.
6
Set the vertical Scale to the Volts/div value from the first row of the table. Set the Offset
7
to the Position value from the first row of the table. Click Close.
µ
F
Supply
Tolerance
±36 mV
2.00 V
±18 mV
1.00 V
±9 mV
500 mV
Chapter 3: Testing Performance

To test offset accuracy

Recommended Model/Part
Agilent 6114A
Agilent 34401A
Agilent 10503A
Agilent 1251-2277
Agilent 1250-0781
Agilent 10240B
Agilent 1250-0774
Limits
minimum
maximum
1.964 V
2.036 V
0.982 V
1.018 V
491 mV
509 mV
3–11