Setup > 3-Port (Ports 1, 2, and 4)
With a 4-port calibration, it is possible to make 3-port measurements. This reduces the
number of sweeps from 8 (for a 4-port device) to 6. Test ports 1, 2, and 4 are used. Test port 3
is not used.
Select Setup > 3-Port (Ports 1, 2, and 4) to change the configuration from 4-Port to 3-Port.
The Test-Set Connections box opens to show the proper connections:
This connection diagram shows the test-set to DUT connections for single-ended
measurements. Click OK to close the Test Set Connections box.
The S-Parameter screen changes from 16 plots to 9 plots. Transforms are disabled whenever
the port configuration is changed.