Menu Reference
Setup Menu
The errors in the unmeasured paths of Config 1 and Config 2 are calculated. Config 3
measures all paths to ensure maximum accuracy when needed.
When a low-loss Thru adapter with a known delay is being used, select Phase Rotation,
enter the delay value, and select the units in picoseconds, pS, or centimeters-in-air, cm (air).
Click Select Adapter... to add adapter (.txs) files to the drop-down selection box. To select
more than one adapter file, hold down the CTRL key while clicking each file. This allows
multiple selections in the drop-down box.
Click Clear All Positions to remove the adapters files from all transmission paths.
Click Use In All Valid Positions to apply the adapter file in all valid positions. The software
calculates this from the connector from the calibration kits selected in Setup > Cal Kit
Selection... (see
page
112).
Click the check boxes near the connector depictions to apply or remove an adapter
characterization file to a transmission path.
If you try to apply an adapter file to a position where it cannot be used. The system will warn
you:
Remember that cable connectors are defined by the calibration kit assigned to them.
Click Interpolate data if necessary to have the software interpolate the adapter
characterization file, if needed, relative to any measurement frequency settings. (This setting
is recommended.)
Click OK to accept the adapter settings.
Click Cancel to exit the dialog box without making any changes.
114
Chapter 5