Setup > De-Embedding...
Pull down Setup > De-Embedding... to open the De-Embedding dialog box and configure
DUT reference planes by de-embedding probes, fixtures, or adapters from the measured data.
The software supports reference plane rotation and S-parameter de-embedding of fixtures.
Select the reference plane rotation using one of two methods:
• Click Rotate Reference Plane and enter a reference plane rotation value in
centimeters-in-air.
• Click Select De-Embed Files to add de-embedding (.txs) files to the drop-down
selection box to use S-parameter de-embedding. You can select more than one fixture (or
adapter characterization) file at once to allow multiple selections in the drop-down box.
Select an S-parameter de-embedding file from the list then click the boxes near the DUT
diagram to apply the de-embed file to the DUT port.
De-embedding files contain information about the ports on the fixture that you are
de-embedding, and match the connectors during de-embed setup. If you try to specify a fixture
connection that cannot be made, the software warns you: