Menu Reference
Setup Menu
Click Copy marker readout to print area to have the system default to include marker
information in the annotation area of the printed copy.
Click the Load/Save tab to display the Load/Save preference options.
Performing a measurement creates a large amount of data in addition to the displayed data.
This includes raw uncorrected data and data generated during intermediate stages of the
correction process.
If de-embedding or alternate port impedance is active, more intermediate data is created.
When you save data using File > Save > Data, the raw data and most of the intermediate
data is removed to reduce the size of the disk file. As a result, certain operations cannot be
reversed if the file is loaded back into the application.
Click Save reference-plane data if data has been de-embedded to allow the removal of
de-embedding information from a data file loaded to a disk. (This is the default setting.)
If Save reference-plane data if data has been de-embedded is not checked, the plots
with the de-embedding information are saved and the raw data used to create the loaded data
file can no longer be viewed. However, information about the de-embedding used on the raw
data is still available for review.
Click Save raw and intermediate data to save all the data. This is only recommended
when you need data for a service call to Agilent. For normal operation this selection is not
recommended because it results in large disk files.
Setup > 4-Port
Select Setup > 4-Port to change the measurement configuration from 3-port or 2-port to the
default 4-port configuration.