Agilent Technologies N4413A User Manual page 136

Multiport test system
Table of Contents

Advertisement

Menu Reference
Setup Menu
Select Perform Setup sequence when Calibrate is selected to force complete calibration
configuration each time you select Calibrate > Perform Calibration > from the main menu.
With Perform Setup sequence when Calibrate is selected enabled, the Measurement
Parameters, Calibration Kit Selection, and CALIBRATION: THRU ADAPTER/PHASE
ROTATION dialog boxes are displayed for edits or confirmation. With this option off, the
system begins calibrating immediately with the current calibration settings.
Select Display all setup information before entering calibration to have the system
show you the current calibration settings before starting the calibration. This allows you to
confirm that the calibration settings are correct before starting.
Verify before exiting incomplete calibration is not implemented.
Click the Measurement tab to display measurement preference options.
126